Ion Contamination Tester AC100
AC100 is an ionic contamination tester for PCBA, PCB and semiconductor parts.
According to the method of ionic contamination degree test defined in IPC-TM-650, the test process includes 3 processes: solvent regeneration, ion contamination solvent extraction and ion contamination degree measurement.
This tester adopts a high-purity, high-capacity ion exchange device. Before each test, the solvent is regenerated through the ion exchange device according to the setting value to ensure that the ion content in the solvent meets the test requirements before each test.
Machine Application:
AC100 is an ionic contamination tester for PCBA, PCB and semiconductor parts.
According to the method of ionic contamination degree test defined in IPC-TM-650, the test process includes 3 processes: solvent regeneration, ion contamination solvent extraction and ion contamination degree measurement.
This tester adopts a high-purity, high-capacity ion exchange device. Before each test, the solvent is regenerated through the ion exchange device according to the setting value to ensure that the ion content in the solvent meets the test requirements before each test.
Machine Features:
-
Using high-precision data collection and detecting measurement system, no solvent polarization phenomenon, to ensure the accuracy of the test.
- Automatic calibration function, no need for complicated calculations.
- Special regeneration device, fast regeneration and long service life.
- Automatically adjust the liquid level according to the size of the PCBA/PCB, both large and small boards can be tested.
- The size of the test tankcan be customized,
- Fully enclosed test environment to reduce the influence of CO2 on the test results.
- Immersion spray wash PCBA/PCBto ensure complete extraction of ionic contamination.
- The operating software is simple and intuitive. The testing process is fully automatic.
- Fully compatible with international standards MIL-P-28809, MIL-STD-2000A, DEF-STD-10/03, etc.
Technical Spec:
S/N |
Item |
Technical Spec |
Remarks |
1 |
Test solvent |
75% IPA/DI water |
|
2 |
Test method |
Static |
|
3 |
Test Range |
0.01-20.00ug NaCl Eq/cm2 |
|
4 |
Resolution |
0.001uS/cm |
|
5 |
Solvent capacity |
10L |
|
6 |
Test accuracy |
±5% |
|
7 |
PCBA size |
L300*W280*H45(mm) |
|
8 |
Machine size |
L620*W380*H600(mm) |
|
9 |
Machine weight |
40Kgs |
|
10 |